Brooks Instrument SolidSense II Integrated-Display Pressure Transducer
About this part
Brooks Instrument GIDX2C7YCSL is a SolidSense II GI-series dead-end pressure transducer for high-purity and ultra-high-purity process-gas monitoring. Its ordering code specifies an integrated display, 235 psi compound range, 2–10 Vdc output, 18-inch pigtail with four-pin AMP connector, and standard 1.125-inch C-Seal surface-mount process connection
Configurations vary. Verify the exact unit before choosing a replacement.
ManufacturerBrooks Instrument SolidSense II GF/GI Series
Part numberGIDX2C7YCSL
Product familyConfiguration-specific
Device typePressure Sensor
Configuration statusConfiguration varies
Application classSemiconductor/high-purity gas delivery
CategoryPressure And Vacuum
Specifications
Identification
Manufacturer
Brooks Instrument
Product family
SolidSense II GF/GI Series
Configured model
GIDX2C7YCSL
Configuration
Body type
Dead end
Display
Integrated, continuously rotatable red LED display
Pressure range
235 psi compound
Analog output
2–10 Vdc
Electrical connection
18-inch pigtail with four-pin AMP connector
Process connection
Standard surface-mount 1.125-inch C-Seal
Electrical
Supply voltage
13–30 Vdc
Typical load resistance
10 kΩ for the 2–10 Vdc output
Performance
Accuracy
±0.25% of full scale, BFSL
Response time
Less than 5 ms
Proof pressure
200% of full scale for this range
Burst pressure
400% of full scale
Environmental
Operating temperature
−20°F to 140°F (−29°C to 60°C), GI display version
Storage temperature
−40°F to 167°F (−40°C to 75°C), GI display version
Materials
Wetted parts
316L stainless steel meeting SEMI F20
Surface finish
Compliant with SEMI F19
Cleanliness
Compliant with ASTM F1374-92 (2005)
Mechanical
Internal volume
1.79 cc
Display
Display accuracy
±0.25% of reading ±1 count for psi, excluding transducer-output accuracy
Unit selection
Psi/kPa switch
Zero adjustment
Provided
What Does GIDX2C7YCSL Mean?
The identifier follows the Brooks SolidSense II product-description code. The final L appears in Brooks CAD configuration names but is not defined by the current public ordering tables.
Segment
Field
Interpretation
GI
Base model
Pressure transducer with integrated display
D
Body type
Dead end
X2
Pressure range
235
C
Pressure units and reference
Compound, psi
7
Output
2–10 Vdc
Y
Electrical connection
18-inch pigtail with four-pin AMP connector
CS
Process fitting
Standard surface-mount 1.125-inch C-Seal
L
Terminal option
Manufacturer CAD configuration suffix; its meaning is not defined in the current public model-code tables
Four-Pin AMP Connection
The manufacturer Type Y voltage-output wiring table assigns the four connector conductors as follows.