Brooks Granville-Phillips / 275806-EU Series 275 Mini-Convectron Vacuum Gauge Module
About this part
T198X32 is an archived unit identifier associated with the Brooks-era Granville-Phillips catalog number 275806-EU. That configuration is a Series 275 Mini-Convectron vacuum gauge module with nonlinear analog output, one setpoint relay, no digital display, and an NW16KF flange. It measures air or nitrogen from 1 × 10^-4 to 1000 Torr and is factory calibrated for nitrogen; measurements of other gases require appropriate conversion data
Configurations vary. Verify the exact unit before choosing a replacement.
T198X32 is not defined in the Granville-Phillips ordering tables. The associated 275806-EU catalog number supplies the documented product configuration.
Segment
Field
Interpretation
T198X32
Identifier type
Archived unit identifier; not a published Granville-Phillips ordering code and therefore not technically decodable.
275806-EU
Observed catalog number
Legacy Series 275 Mini-Convectron module with nonlinear analog output, one setpoint relay, no display, and an NW16KF flange.
EU
Catalog suffix
Included in the legacy CE-marked catalog number; the manual does not publish a separate character-by-character definition for this suffix.
Legacy and Current Configuration Comparison
MKS currently catalogs 275400-1-GD-T with the same principal function set as legacy 275806-EU. A direct drop-in replacement relationship is not established by the manufacturer material cited here.
Field
Legacy 275806-EU
Current 275400-1-GD-T
Interface
Nonlinear analog
Nonlinear analog
Display
None
None
Setpoint relays
One
One
Vacuum fitting
NW16KF
NW16 ISO-KF
Filament
Standard tungsten or optional platinum; installed option unresolved
Gold-plated tungsten
Internal volume
35 cm³
40 cm³
Direct replacement status
Not established in cited manufacturer material
Obtain MKS confirmation before substitution
Gas and Installation Warnings
Do not use the module to measure flammable or explosive gases.
Pressure indication is gas-dependent. Apply manufacturer conversion data for gases other than nitrogen or air, and establish a calibration curve with a gas-independent transfer standard when suitable data are unavailable.
Configure the relay setpoint for the actual process gas before connecting it to system-control equipment.
Maintain the gauge axis horizontally for correct readings above 1 Torr.
Provide pressure-relief protection where system pressure can exceed the instrument limit.