Brooks Instrument SolidSense II GFD 100 psia 4–20 mA Pressure Transducer
About this part
GFD01A4QSM is a non-display, dead-end SolidSense II GF Series transducer configured for 100 psia full scale, a 4–20 mA output, a 12-inch pigtail terminating in an M12 connector, and a 1/4-inch swivel-male face-seal process connection. The GF Series is designed for stable pressure monitoring in high-purity and ultra-high-purity process-gas systems
Configurations vary. Verify the exact unit before choosing a replacement.
ManufacturerBrooks Instrument SolidSense II GF Series
Part numberGFD01A4QSM
Product familyConfiguration-specific
Device typePressure Sensor
Configuration statusConfiguration varies
Application classSemiconductor/high-purity gas delivery
CategoryPressure Transducers
Specifications
Identification
Manufacturer
Brooks Instrument, LLC
Product series
SolidSense II GF Series
Model code
GFD01A4QSM
Configuration
Instrument type
Pressure transducer without integrated display
Body type
Dead-end
Measurement
Full-scale range
100 psia
Pressure reference
Absolute
Electrical
Output signal
4–20 mA
Power requirement
10–30 Vdc
Connection
12-inch pigtail with M12 connector
Process connection
Fitting
1/4-inch face seal, swivel male
Performance
Accuracy
±0.25% of full scale (BFSL)
Response time
Less than 5 ms
Pressure limits
Proof pressure
200% of full scale (200 psi for this range)
Burst pressure
400% of full scale (400 psi for this range)
Environmental
Operating temperature
−20°F to 180°F (−29°C to 82°C)
Storage temperature
−40°F to 180°F (−40°C to 82°C)
Compensated temperature
−4°F to 140°F (−20°C to 60°C)
Construction
Wetted parts
316L stainless steel, SEMI F20
Housing
Stainless steel and polymer plastics
Surface finish
Compliant with SEMI F19
Cleanliness
Compliant with ASTM F1374-92 (2005)
Mechanical
Internal volume
1.79 cc
Approximate shipping weight
0.70 lb (0.32 kg)
Calibration
Factory calibration
Factory calibrated; field zero adjustment is not normally required
What Does GFD01A4QSM Mean?
Brooks' seven-field Product Description Code defines the configured instrument.
Segment
Field
Interpretation
GF
Base model
Pressure transducer without an integrated display
D
Body type
Dead-end
01
Pressure range
100 full scale
A
Pressure units and reference
Absolute pressure in psi; combined with 01, the range is 100 psia
4
Output
4–20 mA
Q
Electrical connection
12-inch pigtail with M12 connector
SM
Process fitting
1/4-inch face seal, swivel male
Type Q M12 Current-Loop Wiring
For the configured 4–20 mA output and Type Q connection, the Brooks wiring table assigns the current-loop supply to M12 pins 1 and 3.
Pin
Assignment
1
+ Supply (+E)
3
− Supply (−E)
Other pins
Not connected
Approval and Installation Notes
Some GFD and GFF configurations are FM approved, but GFD01A4QSM does not establish an approval through its ordering code. Verify the physical product label and applicable certificate before hazardous-area use.
The instrument is factory calibrated and normally requires no field zero adjustment.
Do not use the housing or cable nut as leverage when tightening the process fitting; use a stationary point on the mating gas-line fitting.