Flow Control

5850S/BC1AA1AC3BA0A1

Brooks Instrument Smart-Series Digital Mass Flow Controller

About this part

5850S/BC1AA1AC3BA0A1 is a configured Brooks Instrument 5850S Smart-Series thermal mass flow controller identifier observed in stored Sonisys auction records. Official Brooks 5800S documentation covers the 5850S controller model in the Smart Mass Flow family, but no verified official Brooks PDF or product page is currently available that names the full configured identifier. Exact gas, full-scale flow, calibration gas, fittings, valve state, input/output option, and seal package should be verified from the physical unit label or calibration paperwork

Configurations vary. Verify the exact unit before choosing a replacement.

Key Facts

ManufacturerBrooks Instrument
Part number5850S/BC1AA1AC3BA0A1
Product family5850S
Device typeMass Flow Controller
Configuration statusConfiguration varies
Application classSemiconductor/high-purity gas delivery
CategoryFlow Control

Specifications

Identification
Device typeSmart-Series digital thermal mass flow controller
Series family5800S Smart Mass Flow Series
Configured code5850S/BC1AA1AC3BA0A1
Official covered controller modelsBrooks 5800S documentation covers Smart Mass Flow Controller models 5850S, 5851S, and 5853S
Configuration
Configured-unit ruleGas, full-scale range, fitting, valve state, connector, analog/digital I/O package, and seal package vary by configured model code and unit calibration
Flow
5850S family flow range contextBrooks 5800S documentation lists the 5850S controller family from 0.003 to 30 ln/min nitrogen-equivalent full scale, depending on configuration
Performance
Accuracy+/-0.7% of rate and +/-0.2% full scale at calibration conditions; optional +/-0.5% of rate and +/-0.1% full scale depending on configuration
Repeatability+/-0.25% of rate
Rangeability50:1 within specified accuracy
Controllability100:1 total operating range
Controller settling timeStandard 1 second settling time for 5850S / 5851S controllers, with response up to 0.2 seconds available by configuration
Pressure
Pressure rating contextBrooks 5800S documentation lists 5850S / 5860S pressure rating context of 100/300 bar, with high-pressure operation dependent on configuration
Environmental
Temperature rangeAmbient and process gas: 0 C to 70 C; optional 0 C to 100 C configuration context is listed in Brooks family documentation
Materials
Wetted materialsStainless steel wetted parts with Viton, Buna-N, PTFE/Kalrez, EPDM, or mechanical elastomers depending on configured build
Mechanical
Process connection contextBrooks 5800S documentation lists NPT(F), tube compression, VCR/VCO, downport, and DIN/ANSI flanged connection contexts depending on configuration
Electrical connector15-pin D-type connector with gold-plated contacts; 9-pin D-type connector applies when Profibus-DP is configured
Electrical
Setpoint input0-5 Vdc or 1-5 Vdc voltage input, or 0-20 mA / 4-20 mA current input depending on configuration
Analog output0-5 Vdc, 1-5 Vdc, 0-20 mA, or 4-20 mA flow output depending on configuration
Controller power+15 to +28 Vdc, with -15 Vdc used for selected configurations such as normally open valve builds
5850S controller current drawAt +24 Vdc, 140 mA typical for 5850S / 5851S / 5853S controllers; at +15 Vdc, 185 mA typical
Controls
Digital communicationRS-232 or RS-485 communication with selectable baud rates from 1200 to 38K4; Profibus-DP up to 12 Mbit/sec when configured
Calibration storageDigital communication can access calibration data, including storage of up to 10 calibration curves
DiagnosticsSelf-diagnostics can activate an alarm output and expose failure details through digital communication
Applications
Application domainSemiconductor, laboratory, analytical, and process gas handling
Observed variants
Stored auction gas/range combinationsNo reliable gas or full-scale flow range was extracted from stored auction titles for this exact configured identifier

Documentation Downloads

Local PDF copies

Part Photos

5 images

Historical archive images for identification, labels, connectors, and condition reference, not current inventory.

Historical Listings & Prices

Historical listing data, not a current offer
Archive records2Tracked historical listing records
Historical median price$299.00Based on archived listing records
Recorded price range$299.00 - $299.00Low to high recorded listing price
Local documents2PDF copies available from Sonisys
Recorded price context

The typical recorded price range is $299.00 to $299.00. Based on 2 archived listing records; not a current offer, appraisal, or confirmed sale price.

View archived listing titles
  • Brooks 5850S/BC1AA1AC3BA0A1 Smart Mass Flow 30 Days Warranty Fast Shipping

Archived listings

Other Part Numbers & Aliases

5850S/BC1AA1AC3BA0A15850S BC1AA1AC3BA0A15850SBC1AA1AC3BA0A1Brooks 5850S/BC1AA1AC3BA0A1Brooks Instrument 5850S/BC1AA1AC3BA0A1