Brooks Instrument Series Elastomer-Sealed Gas Mass Flow Controllers and Meters
About this part
GF40 is Brooks Instrument’s series designation for GF040 elastomer-sealed digital thermal gas mass flow controllers and meters. MultiFlo-capable configurations can be reconfigured for supported gases and full-scale ranges. The bare GF40 identifier does not define the gas, calibrated range, valve, fitting, seal compound, downstream condition, or communications interface; those details require the complete configured model code or nameplate
Configurations vary. Verify the exact unit before choosing a replacement.
Viton, Buna, Kalrez, or EPDM; exact compound is configuration-specific
Interfaces
Documented communications options
RS-485, Profibus, DeviceNet, and EtherCAT, with configuration-dependent analog I/O
Compliance
Documented compliance
CE to EN 61326:2006, safe-area design to EN 61010-1, and RoHS
What Does GF40 Mean?
GF40 identifies the product series, while GF040 is Brooks’ documented base-model code. Neither value is a complete configured ordering code.
Segment
Field
Interpretation
GF40
Series identifier
Brooks Instrument GF40 Series of elastomer-sealed gas mass flow controllers and meters.
GF040
Base-model code
Elastomer-sealed base platform covering the documented range through 50 slpm.
Not supplied
Configuration fields
The bare identifier provides no codes for configurability, special application, valve, gas/range, fitting, downstream condition, seals, communications, customer special request, auto shutoff, or reference temperature.
Configuration Identification
Use the complete model code and calibration/nameplate data to establish a unit’s gas, full-scale range, valve arrangement, process fitting, downstream-pressure condition, elastomer compound, electrical interface, customer-specific option, auto-shutoff selection, and reference temperature.
Valve and Response Options
Controller response depends on the ordered valve arrangement.
Valve configuration
Ordering-code value
Documented settling time
Normally closed
C
Less than 1 second
Normally open
O
Less than 3 seconds
Meter, no valve
M
Measure-only configuration; controller settling-time claim does not apply
Differential Pressure by Flow Band
The data sheet specifies configuration-dependent differential-pressure ranges; high-density gases such as argon typically require an additional 10 psid.